Mean kinetic temperature calculator (MKT)

Free mean kinetic temperature calculator: the MKT of a storage or transport period from its temperature readings, with ΔH 83.144 kJ/mol by default.

This mean kinetic temperature calculator weights each reading by the Arrhenius equation, with ΔH ÷ R = 10,000 K by default: readings of 20 and 30 °C give an MKT of 26.3 °C, above their 25 °C average.

Readings at equal intervals, or interval averages, separated by spaces, lines or semicolons.

83.144 kJ/mol is the usual default when the product’s own value is unknown.

Mean kinetic temperature
– °C
Arithmetic mean
– °C
Lowest
– °C
Highest
– °C
Readings
–

MKT weighs warm readings more than the plain mean, because degradation speeds up with temperature.

How it works

How is mean kinetic temperature calculated?

MKT = (ΔH ÷ R) ÷ −ln( (e−ΔH/RT₁ + … + e−ΔH/RTₙ) ÷ n ), with temperatures in kelvin and R = 8.3144 J/mol·K. With 83.144 kJ/mol, ΔH ÷ R is exactly 10,000 K.

It is the single steady temperature that would cause the same chemical degradation as the varying temperatures you entered. It is always at or above the arithmetic mean.

Use readings at equal intervals. MKT summarises a period; it does not excuse an excursion against the product’s labelled limits.

In Hermetry Hermetry keeps daily minimum and maximum temperatures per storage unit from its day aggregates: Medicine storage temperature monitoring.

Glossary Data logger

Questions

Questions about the mean kinetic temperature calculator

Why is MKT higher than the average temperature?

Reaction rates rise exponentially with temperature, so a warm hour does more harm than a cool hour does good. MKT weights each reading by its effect.

Which activation energy should I use?

The product’s own value when known. 83.144 kJ/mol is the usual default.